Paid
SERVICE DESCRIPTION | Instrument : C230 Oxygen Transmission Rate Test System, Brand : Labthink-China C230 Oxygen Transmission Rate Test System is designed and manufactured based on coulometric sensor method (aka. equal pressure method) and conforms to the requirements…
- Laboratorium Karakterisasi Lanjut Kimia Maju I
- KST Serpong (Bacharuddin Jusuf Habibie)
Gedung 456 KST BJ Habibie Tangerang Selatan, Banten , 15314 - 08119811561
- labkarserpong@brin.go.id
Marketing Office
Deputi Bidang Infrastruktur Riset dan Inovasi BRIN
layanan_sains@brin.go.id
Instrument :
C230 Oxygen Transmission Rate Test System, Brand : Labthink-China
C230 Oxygen Transmission Rate Test System is designed and manufactured based on coulometric sensor method (aka. equal pressure method) and conforms to the requirements of ISO 15105-2/ASTM D3985. This instrument can be used to measure the oxygen transmission rate of barrier materials with high, medium and lower barrier properties with a wide testing range and high testing efficiency. The instrument is featured with patented design of integrated block consisting of 3 test cells. Equipped with high precision sensors and professional computer-controlled system, the instrument can regulate and control the temperature, humidity and flow rate properly, which guarantee the testing sensitivity and repeatability of test results. C230 is applicable to determination of oxygen permeability of plastic films, sheeting, paper, packages and other relative packaging materials in food, pharmaceutical, medical apparatus, daily chemical, photovoltaic and electronic industries, etc.
Test Principle
The pre-conditioned specimen is mounted between the upper and lower chambers at ambient atmospheric ressure. One chamber contains oxygen or air and the other chamber is slowly purged by a stream of nitrogen. Due to the concentration difference between the two chambers, oxygen molecules permeate through the specimen into the nitrogen side and are taken to the coulometric sensor where proportional electrical signals are generated. The oxygen transmission rate is then obtained by analyzing and calculating the signals. For package samples, high purity nitrogen flows inside the package, and oxygen or air flows outside.
This test instrument conforms to the following standards:
ASTM D3985
Basic Applications
1. Films
Plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many
others
2. Sheeting
PP, PVC, PVDC, metal foils, rubber pads, silicon wafers and other sheeting materials
3. Packages
Bottles
Specifications :
For more information, please contact us
https://wa.me/628119811561
Application Requirements:
C230 Oxygen Transmission Rate Test System, Brand : Labthink-China
C230 Oxygen Transmission Rate Test System is designed and manufactured based on coulometric sensor method (aka. equal pressure method) and conforms to the requirements of ISO 15105-2/ASTM D3985. This instrument can be used to measure the oxygen transmission rate of barrier materials with high, medium and lower barrier properties with a wide testing range and high testing efficiency. The instrument is featured with patented design of integrated block consisting of 3 test cells. Equipped with high precision sensors and professional computer-controlled system, the instrument can regulate and control the temperature, humidity and flow rate properly, which guarantee the testing sensitivity and repeatability of test results. C230 is applicable to determination of oxygen permeability of plastic films, sheeting, paper, packages and other relative packaging materials in food, pharmaceutical, medical apparatus, daily chemical, photovoltaic and electronic industries, etc.
Test Principle
The pre-conditioned specimen is mounted between the upper and lower chambers at ambient atmospheric ressure. One chamber contains oxygen or air and the other chamber is slowly purged by a stream of nitrogen. Due to the concentration difference between the two chambers, oxygen molecules permeate through the specimen into the nitrogen side and are taken to the coulometric sensor where proportional electrical signals are generated. The oxygen transmission rate is then obtained by analyzing and calculating the signals. For package samples, high purity nitrogen flows inside the package, and oxygen or air flows outside.
This test instrument conforms to the following standards:
ASTM D3985
Basic Applications
1. Films
Plastic films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films, glass fiber aluminum foil composite films and many
others
2. Sheeting
PP, PVC, PVDC, metal foils, rubber pads, silicon wafers and other sheeting materials
3. Packages
Bottles
Specifications :
- Method : Coulometric sensor method (aka. equal pressure method)
- Test Range : (Standard) 0.05~5000 cm3/(m2d·)
- Test Chamber : 3 test chambers
- Repeatability : 0.0025 and 2%
- Resolution : 0.001 g/(m2d·)
- Test Temperature : 10~55 ±0.2 ℃
- Test Humidity : 0%, 5% ~ 90%±1%RH
- Any samples that have been tested and do not produce results are still subject to normal rates and analysis fees are non-refundable
- For film samples the minimum sample diameter is 13 cm; Maximum sample thickness is 3 mm
- Samples that have been tested will be stored in the laboratory for a maximum of 3 months, then they will be destroyed
- Maximum payment time is one month (1 month) after the Billing Code is issued
For more information, please contact us
https://wa.me/628119811561
Application Requirements:
- Photo
- Other Supporting File
File Name | File Size | |
---|---|---|
Service SOP file | 0.44 MB | |
Template Photo | 0 MB | |
Template Other Supporting File | 0 MB |